The Ai DeepFreeze series -40C freezers is perfect for large storage of sample and products. Our unique single-compressor design offers low power consumption, exceptionally low noise, and ease of maintenance.
Features
- Great for storage of Covid-19 vaccines, DNA, RNA, virus samples, organs and serum
- High/low temperature alarms, power failure alarm, door open alarm, filter clogging alarm, password protection, tons of security features to protect you, your products, and the unit
- Dual silicon sealing gaskets for ultra airtight operation
- 48-hour battery backup for controller
- Built-in remote alarm and RS-485 port
- USB interface for data export and storage
- Environmentally friendly refrigerant
- Use only RoSH compliant non-toxic components and materials
Important!
- Do not plug in your unit. Leave your Ai refrigerator or freezer for 24 hours before turning the unit on for the first time.
- This unit is NOT designed to be used on a mobile platform.
Model | D16 |
Electrical requirements |
110V (±10%) 60Hz 1-Phase 8.89A |
Performance | Temperature range: -20°C to -40°C |
Compressor |
Number of compressors: two Cooling method: direct cooling Defrost mode: manual Style: parallel Refrigerant: R290 |
Chamber | Dimensions: 25.5 x 22.5 x 24.5" (WxDxH) Number of chambers: two Material: stainless steel Volume: 15.9 Cu Ft |
Shelf | Quantity: two in each chamber, four in total Style: fixed Material: stainless steel |
Door |
2 |
Working environment | - Temperature: 16°C to 32°C - Humidity: <80% Rh - Altitude: < 2000 m / 6561 Ft - Stable, well-ventilated platform with no excessive dust - Keep freezer at least 12 inches from any walls or ceiling - Use dedicated power, do not share outlet - Extension power cable should be less than 10 feet is required - Freezer and power outlet should be grounded properly - Avoid direct sunlight, water spill and do NOT use freezer outdoor - Clean condenser fins and remove frost inside chamber at least once a month - Important! Do not plug in your unit. Leave your Ai refrigerator or freezer for 24 hours before turning the unit on for the first time. |
Access port | Two |
Communication | Remote alarm 4-20mA transmitter, USB, RS-485 |
Built-in swivel casters | Four, two with brakes |
Noise level |
52 dBA |
Unit/shipping weight & dimensions |
320 Lbs / 345 Lbs 30.5 x 26.5 x 73" / 37.5 x 34.5 x 80" (WxDxH) |
Safety | Compressor over-temp protection Compressor power surge delay protection Power failure protection Low battery, door open, filter clogging, system failure, sensor error, condenser overheat alarms External door with lock |
Warranty | One year |